<p id="jg8hh"></p><delect id="jg8hh"><em id="jg8hh"><blockquote id="jg8hh"></blockquote></em></delect><acronym id="jg8hh"><dd id="jg8hh"></dd></acronym><button id="jg8hh"><dd id="jg8hh"><acronym id="jg8hh"></acronym></dd></button><samp id="jg8hh"><em id="jg8hh"><blockquote id="jg8hh"></blockquote></em></samp>

<p id="jg8hh"></p>

<samp id="jg8hh"><legend id="jg8hh"></legend></samp>
<samp id="jg8hh"><legend id="jg8hh"><samp id="jg8hh"></samp></legend></samp>

<samp id="jg8hh"></samp>

<p id="jg8hh"></p><acronym id="jg8hh"></acronym><p id="jg8hh"><dd id="jg8hh"><acronym id="jg8hh"></acronym></dd></p><p id="jg8hh"></p>

<p id="jg8hh"></p><delect id="jg8hh"><legend id="jg8hh"><var id="jg8hh"></var></legend></delect><button id="jg8hh"><listing id="jg8hh"><i id="jg8hh"></i></listing></button>
<delect id="jg8hh"><legend id="jg8hh"><var id="jg8hh"></var></legend></delect>

當氣相色譜儀噪聲大的原因

發(fā)布時間:2024-04-24
任何設備或者儀器在工作時都會產(chǎn)生聲音,這個是避免不了的,同樣氣相色譜儀也是如此,尤其是在剛開始工作或者剛更換完色譜柱的時候,其聲音更大,我們一般稱為噪聲。除了剛啟動時或者更換色譜柱后其他比較大的噪聲都是不正常的,這時需要我們排查找出原因進行解決。本文就氣相色譜儀噪聲來進行講解,如何降低氣相色譜儀噪聲。
首先我們分析是否是進樣口有污染,我們需要降低進樣口溫度,如果噪聲降低則考慮進樣口有污染現(xiàn)象;將火焰熄滅之后噪聲如果還是很大,要考慮從檢測器到放大器電路板這一塊找問題,檢查檢測器是否被污染,如果污染請進行清洗,要考慮是極化電壓、放大器電路板、工作電源的故障,檢查檢測器的噴嘴、收集極、離子信號線插座、點火線等部分是否固定可靠,請排除接觸不良的可能;降低色譜柱溫度后如果噪聲變小,要考慮是否載氣純度不夠或色譜柱的老化不足;將火焰熄滅之后噪聲如果降低或消失,要考慮是否檢測器本身產(chǎn)生過大噪聲:檢查空調(diào)器等冷暖設備的排風是否正對著氣相色譜儀,請改變風向或更換儀器的位置,檢查檢測器是否被污染,如果污染請進行清洗,檢查是否使用的氣體純度太低,請更換氣體或使用氣體過濾器去除氣體中的雜質(zhì);改變量程范圍,噪聲的大小還是基本不變時,要考慮是否信號線的故障、放大器電路板的故障、輸出信號線的故障、積分儀的故障。
上一個:工業(yè)內(nèi)窺鏡如何運用于基礎(chǔ)建設的橋梁
下一個:公積金貸款自建房可以嗎

探討一下壓力校準儀的重要性
生物惰性液相質(zhì)譜聯(lián)用系統(tǒng)提升寡核苷酸定量分析性能
2022年版鼓勵外商投資產(chǎn)業(yè)目錄發(fā)布 含儀器儀表制造業(yè)
電子拉伸試驗機屈服點誤差解決方法
2018年醫(yī)藥行業(yè)將發(fā)生什么變化?
瀏覽器窗口打開多個網(wǎng)頁怎么設置(網(wǎng)頁怎么開多個窗口)
海南承試三級設備清單|海南承裝修試三級資質(zhì)設備清單選型指南
便攜式氣壓計
信百諾30m³空氣凈化器檢測環(huán)境艙讓您遠離霧霾
聯(lián)想y470進入bios設置u盤啟動(聯(lián)想y470u盤啟動快捷鍵)