<p id="jg8hh"></p><delect id="jg8hh"><em id="jg8hh"><blockquote id="jg8hh"></blockquote></em></delect><acronym id="jg8hh"><dd id="jg8hh"></dd></acronym><button id="jg8hh"><dd id="jg8hh"><acronym id="jg8hh"></acronym></dd></button><samp id="jg8hh"><em id="jg8hh"><blockquote id="jg8hh"></blockquote></em></samp>

<p id="jg8hh"></p>

<samp id="jg8hh"><legend id="jg8hh"></legend></samp>
<samp id="jg8hh"><legend id="jg8hh"><samp id="jg8hh"></samp></legend></samp>

<samp id="jg8hh"></samp>

<p id="jg8hh"></p><acronym id="jg8hh"></acronym><p id="jg8hh"><dd id="jg8hh"><acronym id="jg8hh"></acronym></dd></p><p id="jg8hh"></p>

<p id="jg8hh"></p><delect id="jg8hh"><legend id="jg8hh"><var id="jg8hh"></var></legend></delect><button id="jg8hh"><listing id="jg8hh"><i id="jg8hh"></i></listing></button>
<delect id="jg8hh"><legend id="jg8hh"><var id="jg8hh"></var></legend></delect>

原子熒光光度計原子熒光空白值過低怎么辦

發(fā)布時間:2024-10-01
原子熒光光度計原子熒光空白值過低怎么辦
原子熒光光度計原理是理化實驗室常用的元素分析設備,原理是利用或作為還原劑,將樣品溶液中的待分析元素還原為揮發(fā)性共價氣態(tài)氫化物(或原子蒸汽),然后借助載氣將其導入原子化器,在氬-氫火焰中原子化而形成基態(tài)原子,通過測量待測元素的原子蒸氣在輻射能激發(fā)下產生的熒光發(fā)射強度,來確定待測元素含量的方法。
一、聯機故障。原因是主機先開而電腦后開,造成主機和電腦通訊故障,無法聯機,解決辦法是關掉主機電源之后再開主機電源。
二、原子熒光光度計原理原子熒光空白值過低。在走載流平衡儀器時熒光值明顯低于正常值,主要原因有:一是水封沒有水,二是系統漏氣,三是原子化器位置過低,四是電爐絲沒有點火,五是爐芯破損,六是元素燈損壞。解決辦法:一是水封注水;二是系統檢漏;三是調整原子化器高度至合適位置;四是更換破損件。
三、儀器靈敏度不夠。一般來講標液*個點的熒光值應該在100以上,若熒光值低于100測定出來的誤差會較大。解決辦法:一是改變濃度,二是增加燈電流和負高壓,三是改變原子化器高度。
四、儀器長時間無法穩(wěn)定。一般儀器狀態(tài)穩(wěn)定需要30分鐘左右,若超過40分鐘還未穩(wěn)定可視為長時間無法穩(wěn)定。主要原因:一是元素燈損壞,二是載流和還原劑雜質干擾較大,三是儀器設備硬件問題,四是器皿不干凈。解決措施:一是更換元素燈,二是更換試劑藥品品牌,三是請專業(yè)維修工程師上門維修,四是所有器皿均用20%的硝酸浸泡過夜,超聲波清洗。